UNITED STATES
SECURITIES AND EXCHANGE COMMISSION
Washington, D.C. 20549
FORM 8-K
Current Report Pursuant to Section 13 or 15(d) of
the Securities Exchange Act of 1934
Date of Report (Date of earliest event reported): September 16, 2014
SUNSHINE HEART, INC.
(Exact Name of Registrant as Specified in its Charter)
Delaware (State or other jurisdiction of |
|
001-35312 (Commission File No.) |
|
68-0533453 (IRS Employer |
12988 Valley View Road
Eden Prairie, Minnesota 55344
(Address of Principal Executive Offices) (Zip Code)
(952) 345-4200
(Registrants Telephone Number, Including Area Code)
Not Applicable
(Former Name or Former Address, if Changed Since Last Report)
Check the appropriate box below if the Form 8-K filing is intended to simultaneously satisfy the filing obligation of the registrant under any of the following provisions (see General Instruction A.2. below):
o Written communications pursuant to Rule 425 under the Securities Act (17 CFR 230.425)
o Soliciting material pursuant to Rule 14a-12 under the Exchange Act (17 CFR 240.14a-12)
o Pre-commencement communications pursuant to Rule 14d-2(b) under the Exchange Act (17 CFR 240.14d-2(b))
o Pre-commencement communications pursuant to Rule 13e-4(c) under the Exchange Act (17 CFR 240.13e-4(c))
Item 7.01 Regulation FD Disclosure.
On September 16, 2014, Sunshine Heart, Inc. (Sunshine Heart or the Company) announced that follow-up data from six patients implanted with the C-Pulse® Heart Assist System at St. Lukes Hospital Mid America Heart Institute, as part of a FDA approved, North American Feasibility Trial, will presented by Dr. Sanjeev Aggarwal, MD, Director of Mechanical Circulatory Support at Saint Lukes Mid America Heart Institute. Prior to his clinical presentation at TCT Conference, Dr. Aggarwal will present these case studies and a COUNTER HF study update to the investment community at Sunshine Hearts fourth annual analyst and investor breakfast that begins at 7:00am (EDT) at the Grand Hyatt Washington on September 16, 2014. Dr. Christopher Bowles, PhD, Royal Brompton and Harefield NHS Foundation Trust and Imperial College London, and Dr. William E. Cohn, MD, Texas Heart Institute, will also present at the breakfast. Following the event, the presentations will be available on the investor section of the Sunshine Heart website at http://ir.sunshineheart.com/index.cfm.
Limitation of Incorporation by Reference
In accordance with General Instruction B.2. of Form 8-K, this information including Exhibits 99.1, 99.2, 99.3 and 99.4, is furnished pursuant to Item 7.01 and shall not be deemed to be filed for the purpose of Section 18 of the Securities Exchange Act of 1934, as amended, or otherwise subject to the liabilities of that section. The information in this Item 7.01 of this Current Report on Form 8-K will not be deemed an admission as to the materiality of any information that is required to be disclosed solely by Regulation FD.
Item 9.01 Financial Statements and Exhibits.
Exhibit No. |
|
Description |
99.1 |
|
Press Release - Sunshine Heart to Provide Case Studies from North American Feasibility Trial and Updates on OPTIONS HF and COUNTER HF at 2014 Transcatheter Cardiovascular Therapeutics (TCT) Conference |
99.2 |
|
Presentation - Dr. Sanjeev Aggarwal, Director of Mechanical Circulatory Support at St. Lukes Mid America Heart Institute |
99.3 |
|
Presentation - Christopher Bowles, PhD, Artificial Heart Specialist at Royal Brompton and Harefield NHS Foundation Trust & Imperial College London |
99.4 |
|
Presentation - Dr. William E. Cohn, Director of Minimally Invasive Surgical Technology at Texas Heart Institute at St. Lukes Medical Center |
SIGNATURES
Pursuant to the requirements of the Securities Exchange Act of 1934, the registrant has duly caused this report to be signed on its behalf by the undersigned, hereunto duly authorized.
Dated: September 16, 2014 |
|
SUNSHINE HEART, INC. |
|
|
|
|
By: |
/S/ JEFFREY S. MATHIESEN |
|
Name: |
Jeffrey S. Mathiesen |
|
Title: |
Chief Financial Officer |
EXHIBIT INDEX
Exhibit No. |
|
Description |
99.1 |
|
Press Release - Sunshine Heart to Provide Clinical Updates from the North American Feasibility Study and OPTIONS HF European Post-Market Study at 2014 Transcatheter Cardiovascular Therapeutics (TCT) Conference |
99.2 |
|
Presentation - Dr. Sanjeev Aggarwal, Director of Mechanical Circulatory Support at St. Lukes Mid America Heart Institute |
99.3 |
|
Presentation - Christopher Bowles, PhD, Artificial Heart Specialist at Royal Brompton and Harefield NHS Foundation Trust & Imperial College London |
99.4 |
|
Presentation - Dr. William E. Cohn, Director of Minimally Invasive Surgical Technology at Texas Heart Institute at St. Lukes Medical Center |
Exhibit 99.1
Sunshine Heart to Provide Clinical Updates from the North American Feasibility Study
and OPTIONS HF European Post-Market Study at
2014 Transcatheter Cardiovascular Therapeutics (TCT) Conference
EDEN PRAIRIE, Minn., September 16, 2014 (GLOBE NEWSWIRE) Sunshine Heart, Inc. (Nasdaq: SSH) announced today that follow-up data from six patients implanted with the C-Pulse® Heart Assist System at St. Lukes Hospital Mid America Heart Institute, as part of a FDA approved, North American Feasibility Study, will be presented later today by Dr. Sanjeev Aggarwal, MD, Director of Mechanical Circulatory Support at Saint Lukes Mid America Heart Institute. Prior to his clinical presentation at the TCT Conference, Dr. Aggarwal will present on a few of these patients who demonstrated clinical signs of myocardial recovery to the investment community at Sunshine Hearts fourth annual analyst and investor breakfast that begins at 7:00am (EDT) today at the Grand Hyatt Washington. Dr. Christopher Bowles, PhD, Royal Brompton and Harefield NHS Foundation Trust and Imperial College London, and Dr. William E. Cohn, MD, Texas Heart Institute, will also present at the breakfast. Following the event, the presentations will be available on the investor section of the Sunshine Heart website at: http://ir.sunshineheart.com/index.cfm .
Myocardial Recovery in Ambulatory Heart Failure Patients Treated with the C-Pulse Cardiac Assist System: A Single-Center Experience, will be presented by Dr. Aggarwal in Session II: Mechanical Circulatory Support for Chronic CHF at 4:48 p.m. (EST) in Level 1, Room 152A/B. The patient data outlined on Dr. Aggarwals poster demonstrates that six NYHA Class III patients were successfully implanted with the C-Pulse System, using minimally invasive surgical (MIS) techniques, from July 2010 to April 2012 as part of the North American Feasibility Study. Safety endpoints at 6 months demonstrated that there were no device related deaths, neurological events, aortic disruptions, myocardial infarctions, and mediastinal infections. Quality of life was evaluated using two standard cardiac disease, self-assessments, MLWHF and KCCQ. One patient was transitioned to an implantable LVAD 97 days post C-pulse implantation due to worsening heart failure symptoms. Two patients remain clinically stable on C-Pulse support at 1178 and 982 days respectively, and have showed improvement in NYHA class, 6-minute walk test, and quality of life scores. The other three patients all demonstrated clinically significant improvement allowing for discontinuation of C-Pulse support with explantation of the percutaneous lead and continue to be followed. The mean duration of support was 659 days (range 534-793), the mean ejection fraction of blood pushed from the left ventricle out of the heart increased from 18.3% to 29.3%, and the mean reduction in Left Ventricle End Diastolic Dimension (LVEDD) was 1.1cm. Mean follow-up time post weaning was 335 days (range 52-565).
Dr. Bowles presentation provides a detailed case study of the first patient to be implanted with the C-Pulse System in the UK as part of the OPTIONS HF Study. OPTIONS HF is a post-market, multi-center, perspective, open label study for NYHA Class III and ambulatory Class IV heart failure patients being conducted in the EU. Dr. Bowles will discuss how the first patient qualified for the study, the patients heart failure history and classification, comorbidities, surgical outcome, and current condition.
Dr. Cohns presentation highlights C-Pulse II. C-Pulse II is a totally implantable device that has the same non-blood contacting, non-obligatory and MIS implantation characteristics as the current C-Pulse System alleviating the need for the percutaneous drive line.
About the C-Pulse® Heart Assist System
The C-Pulse Heart Assist System, or C-Pulse System, an investigational device in the United States, Canada and countries that do not recognize the CE mark approval, utilizes the scientific principles of intra-aortic balloon counterpulsation applied in an extra-aortic approach to assist the left ventricle by reducing the workload required to pump blood throughout the body, while increasing blood flow to the coronary arteries. Combined, these potential benefits may help sustain the patients current condition or, in some cases, reverse the heart failure process, thereby potentially preventing the need for later-stage heart failure devices, such as left ventricular assist devices (LVADs), artificial hearts or transplants. It may also provide relief from the symptoms of Class III and ambulatory Class IV heart failure and improve quality of life and cardiac function. Based on the results from our feasibility study, we also believe that some patients treated with our C-Pulse System will be able to be weaned from using the device due to sustained improvement in their heart failure condition as a result of the therapy.
Caution: Investigational device, limited by Federal (or United States) Law to Investigational use.
About Sunshine® Heart
Sunshine Heart, Inc. (NASDAQ: SSH) is an early-stage medical device company focused on developing, manufacturing and commercializing the C-Pulse System for treatment of Class III and ambulatory Class IV heart failure. Sunshine Heart has completed an approved U.S. Food and Drug Administration (FDA) feasibility clinical study of the C-Pulse System and presented the results in November 2011. In March 2012, the FDA notified the Company that it could move forward with an investigational device exemption (IDE) application. Sunshine Heart received unconditional approval from the FDA in November 2012 to initiate its pivotal study. In July 2012, Sunshine Heart received CE Mark approval for its C-Pulse System in Europe. Sunshine Heart is a Delaware corporation headquartered in Minneapolis with a wholly owned subsidiary in Australia. The Company has been listed on the NASDAQ Capital Market since February 2012.
Forward-Looking Statements
Certain statements in this release are forward-looking statements that are based on managements beliefs, assumptions, expectations, and information currently available to management. All statements that address future operating performance, events or developments that we expect or anticipate will occur in the future are forward-looking statements, including, without limitation, future clinical study activities and results including patient enrollment in studies. These forward-looking statements are subject to numerous risks and uncertainties, including, without limitation, the possibility that our clinical studies do not meet their enrollment goals, meet their endpoints or otherwise fail, that regulatory authorities do not accept our application or approve the marketing of the C-Pulse System, the possibility that we may be unable to raise the funds necessary for the development and commercialization of our products, that we may not be able to commercialize our products successfully in the EU and the other risk factors described under the caption Risk Factors and elsewhere in our filings with the SEC. You should not place undue reliance on forward-looking statements because they speak only as of the date when made and may turn out to be inaccurate. We do not assume any obligation to publicly update or revise any forward-looking statements, whether as a result of new information, future events or otherwise. We may not actually achieve the plans, projections or expectations disclosed in forward-looking
statements, and actual results, developments or events could differ materially from those disclosed in the forward-looking statements.
For further information, please contact:
Investor: |
Media: |
Candice Knoll |
David Schull |
Blueprint Life Science Group |
Russo Partners |
T: +1-415-375-3340 |
T: +1-212-845-4271 |
|
|
Jeff Mathiesen |
Andreas Marathovouniotis |
Chief Financial Officer |
Russo Partners |
Sunshine Heart, Inc. |
T: +1-212-845-4235 |
T: +1-952-345-4200 |
|
Exhibit 99.2
SSH_logo_RGB© 2014 Sunshine Heart, Inc. TCT Breakfast Session September 16, 2014 www.sunshineheart.com |
SSH_logo_RGBForward Looking Statement This presentation contains forward-looking statements. All forward-looking statements are managements present expectations of future events and are subject to a number of risks and uncertainties. Various factors could cause actual results to differ materially from these statements including timing, clinical enrollment, clinical results, financing availability, product sales and marketing or efficacy of products, and the other risks set forth under the caption Risk Factors and elsewhere in our periodic and other reports filed with the U.S. Securities and Exchange Commission, including our Annual Report or Form 10-K for the fiscal year ended December 31, 2013. Although the Company believes that the forward-looking statements are reasonable and based on information currently available, it can give no assurances that the Companys expectations are correct. All forward looking statements are expressly qualified in their entirety by this cautionary statement. Caution: C-Pulse ® is an investigational device. The device is limited by federal (United States) law to investigational use only. C-Pulse is a registered trademark of Sunshine Heart Inc. 1 © 2014 Sunshine Heart, Inc. | Slide |
SSH_logo_RGBAgenda Welcome Company Updates and Introductions OPTIONS HF Post Market Multicenter Trial: Early Experience in Europe: a case study Harefield Hospital UK Progress of Fully Implantable Pump Single Center Experience with Patient Weaning from the US Feasibility Trial and thoughts from a leading enroller in the COUNTER HF US Pivotal Study Can C- Pulse Use Lead To Left Ventricular Recovery? Q&A Jeff Mathiesen Dave Rosa Christopher Bowles, PhD Dr. William E. Cohn Dr. Sanjeev Aggarwal © 2014 Sunshine Heart, Inc. 2 |
SSH_logo_RGBIntroductions Dave Rosa © 2014 Sunshine Heart, Inc. |
C-Pulse® System Feasibility Study with a Non-Blood Contacting Extra-Aortic Counterpulsation System in Patients with Moderate to Severe Ambulatory Heart Failure Sanjeev Aggarwal, MD Director, Mechanical Circulatory Support Saint Luke's Mid America Heart Institute Kansas City, MO Caution: Investigational device, limited by Federal (or United States) Law to Investigational use. |
Disclosures No financial disclosures |
C-Pulse Heart Assist System Implantable components: Inflatable balloon placed around the ascending aorta Epicardial ventricular sensing leads Percutaneous driveline for system control and pneumatic actuation |
C-Pulse Heart Assist System Efficacy of counterpulsation therapy Implantation without cardiopulmonary bypass Extravascular location without the need for systemic anticoagulation Untethering Minimally invasive implantation |
Goals of the C-Pulse Feasibility Study A prospective, open label, single arm study Demonstrate feasibility of device and procedure Assess learning curve Refinement of technology and implant technique Safety Efficacy Support conduct of subsequent pivotal trial |
C-Pulse Trial: Primary Safety Endpoints* (6 Months) Death Aortic Disruption Neurologic Dysfunction Myocardial Infarction Major Infection Any other device-related adverse event (as adjudicated by the CEC) *Device related events as defined by INTERMACS event classifications |
C-Pulse Trial: Primary Efficacy Endpoints (6 Months) NYHA Class ranking Quality of Life - Minnesota Living with Heart Failure Score (MLWHF) 6 Minute Walk Test Peak VO2 |
C-Pulse Trial: Other Efficacy Measures (6 Months) Hemodynamics/ RHC (CO/ CI, PAP, PCWP) Quality of Life KCCQ Blood analysis (Na, Cr, Bili, Hgb, LFTs) Concomitant cardiovascular medications Re-hospitalization (HF and all cause) Duration of Support/ survival duration LOS (ICU/ discharge) Device usage/ compliance |
Patient Characteristics N = 20 Mean Age + SD years (range) 56 + 9 (34-71) Gender Female 8 Male 12 Race African American 3 Caucasian 17 NYHA Class Ranking Class III 18 Class IV 2 INTERMACS Classification 3. Stable but inotrope dependent 3 5. Exertion intolerant 8 6. Exertion limited 7 7. Advanced NYHA Class III 2 Entiology Ischemic 8 Non-ischemic 12 |
|
Characteristics of Implant Procedure Numbers are Mean SD (N), Median [Min, Max] for Continuous variables and Percent (Count/N) for discrete variables. Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Fe 2014. In review. Measure All Patients (N=20) Incision to Dressing Time (minutes) 165.7 + 42.4 (19) 156.0 [98.0, 247.0] Anatomical Approach Full Sternotomy 70.0% (14/20) Partial Sternotomy 10.0% (2/20) Right Parasternal 20.0% (4/20) Time in ICU (days) 2.2 + 2.6 (19) 1.1 [0.6, 11.1] Time in Hospital (days) 9.9 + 4.2 (19) 8.0 [4.0, 19.0] |
Device Related Primary Safety Endpoints (6 months and 12 months) Death 0 at 30 days 1 at 6 months Aortic Disruption 1 at 137 days post implant Sternal wound infection (mediastinitis) post surgery CEC adjudicated as major infection, localized, procedure related Infection unresolved despite repeated surgical interventions Aortic rupture occurred at time of third surgical intervention Neurological Dysfunction: None Myocardial Infarction: None Major Infection: 9 patients 8 exit site related 1 PICC line |
Results: NYHA Class 0% 20% 40% 60% 80% 100% I II III IV Baseline N=20 6 Mo N=15 12 Mo N=12 90% 10% 27% 20% 53% 25% 17% 58% p6Mo = 0.0005; p12Mo = 0.0002 Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Feb 2014. In review. |
Results: 6MWT Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Feb 2014. In review. 0 100 200 300 400 500 Meters 276 296 337 p6Mo = 0.2239; p12Mo = 0.0425 Baseline N=20 6 Mo N=15 12 Mo N=12 |
Results: MLWHF Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Feb 2014. In review. 0 10 20 30 40 50 60 70 80 90 Score p6Mo = 0.0005; p12Mo = 0.0005 64.4 40.2 36.4 Baseline N=20 6 Mo N=15 12 Mo N=12 |
Results: KCCQ Overall Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Feb 2014. In review. 0 20 40 60 80 100 Score 41.0 65.6 67.4 p6Mo = 0.0002; p12Mo = 0.0024 Baseline N=20 6 Mo N=15 12 Mo N=12 |
Results: KCCQ Clinical Score Ambulatory Extra-Aortic Counterpulsation in Patients with Moderate to Severe Chronic Heart Failure. Submitted for publication, Feb 2014. In review. 0 20 40 60 80 100 Score 47.9 69.0 73.6 p6Mo = <0.0001; p12Mo = 0.0093 Baseline N=20 6 Mo N=15 12 Mo N=12 |
Worsening HF Hospitalizations Three (3) subjects had worsening heart failure rehospitalizations Two were rehospitalized at 205 and 208 days Did not maintain compliance with device usage One went to LVAD at 4 months Subject ID Date of Implant HF Rehospitalizations Days to HF Event 08-001 21-Jul-10 14-Feb-11 208 08-003 26-Aug-10 18-Mar-11; 06-Aug-11 205 and 345 days 08-005 14-Jan-1 15-Feb-11; 07-Mar-11 33 and 52 days |
Minimally Invasive Surgical Approach |
[LOGO] |
[LOGO] |
Additional Observations 30.9 total patient-years of follow-up Diuretic doses reduced/discontinued in 6 patients Inotropes discontinued in all inotrope-dependent patients 4 patients successfully bridged to transplant 3 permanently discontinued from therapy after sustained improvement |
Myocardial Recovery Three of six patients showed clinically significant improvement allowing for discontinuation of device support with explantation of the percutaneous lead |
Myocardial Recovery Mean duration of support with C-Pulse was 659 days (range 534 793 days) Mean EF improved from 18.3% to 29.3% (one patient EF has improved from 21% to greater than 40%) Mean reduction in LVEDD of 1.1cm No readmissions for heart failure (follow up time range approximately 1-2.5 years) |
Case #1 36 year old male NICM/ chronic systolic heart failure Significant exercise intolerance, DOE Cpulse implant 4/10/2012 Discharged home POD#7 Pre discharge echo demonstrated improvement in EF from 21% to >40% Last use of device 9/25/2013 (533 days of support) PIL Explanted 4/2014 No readmissions for heart failure (>350 days F/U) |
[LOGO] |
Case #2 RA 8 PAP 39/21/26 PCWP 11 CO/ CI 4.3/2.2 RA 7 PAP 47/25/31 PCWP 10 CO/CI 4.2/2.2 |
Conclusions The C-Pulse Study demonstrates the feasibility and shows the preliminary safety and efficacy of the C-Pulse System in patients with moderate to severe heart failure FDA approved the IDE for the COUNTER HF randomized controlled trial to demonstrate the effectiveness of the device |
[LOGO] |
Exhibit 99.3
Options HF C-Pulse System European MultiCentre Study Case Presentation Dr Christopher Bowles Artificial Heart Specialist Dr Nicholas Banner Consultant in Cardiology Transplant and Mechanical Circulatory Support EU: C-Pulse is CE Marked US: Caution: Investigational Device. Limited by Federal (or United States) law to investigational use. 1 |
OPTIONS HF: Key Study Qualifications 2 Major Inclusion Criteria LVEF <35% Both ischemic/non-ischemic NYHA Class III/Ambulatory Class IV On Optimal Medical Therapy Evaluated or have CRT/CRT-D Major Exclusion Criteria Ascending aortic calcification or CABG Mitral Valve Incompetence, Grade 4+ Aortic Valve Incompetence, Grade 2-4+ Confidential © 2014 Sunshine Heart, Inc. |
54 yr old male patient Blood Group: A Rh D Positive Height: 5ft 11inch Wt: 90.6kg Body Mass Index: 28 Confidential © 2014 Sunshine Heart, Inc. 3 |
Background Acute myocardial infarction whilst on vacation in 2002 (required resuscitation following cardiac arrest) Primary percutaneous coronary intervention to right coronary artery (RCA) & left anterior descending (LAD) artery (the latter was unsuccessful resulting in LAD obstruction). Circumflex artery unobstructed Persistent New York Heart Association Class III symptoms since 2002, fatigue and shortness of breath on minimal exertion, loss of employment Atrial fibrillation treated with: - AV node ablation: 1 Nov 2012 - Biventricular ICD pacemaker Type 2 diabetes Advanced heart failure with C-Pulse system intended as a destination to heart transplantation 4 Confidential © 2014 Sunshine Heart, Inc. |
Fulfilled multi-centre study entry criteria C-Pulse implanted at Harefield Hospital, London on 16 June 2014 The first C-Pulse case in the UK 5 Confidential © 2014 Sunshine Heart, Inc. |
Post-operatively Uncomplicated surgical procedure and short Intensive Care Unit stay (28 hours) Increased inflammatory markers after surgery No source of infection identified Given empirical antibiotic therapy Continuing signs of heart failure reflecting poor pre-operative heart function Treated with high dose diuretics. Discharged home on post operative day 22 Readmitted post operative day 47 to 52 with diarrhoea possibly due to antibiotic therapy Readmitted post operative day 64 to 79 with deterioration in exercise capacity attributed to heart rhythm disturbance Treated with short course of i.v. milrinone and higher beta blocker dose 6 Confidential © 2014 Sunshine Heart, Inc. |
Post operative day Distance walked in six min (metre) Heart rhythm disturbance Functional capacity improvement as a result of C-Pulse therapy 7 New ventricular ectopies are an early sign that previously-thought ischemic (or lost) tissue is being perfused Confidential © 2014 Sunshine Heart, Inc. 0 50 100 150 200 250 300 350 400 450 0 20 40 60 80 100 |
Post operative day Distance walked in six min (metre) Heart rhythm disturbance Functional capacity improvement and changes in heart function as a result of C-Pulse therapy Cardiac index Litre/min/m2 With C-Pulse Without C-Pulse Attended daughters marriage 8 Confidential © 2014 Sunshine Heart, Inc. 0 50 100 150 200 250 300 350 400 450 0 20 40 60 80 100 0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 |
Summary 54 year old man Advanced heart failure due to ischaemic cardiomyopathy Uneventful C-Pulse implantation as a bridge to heart transplant Has needed medical management for heart failure since surgery Limited improvement in resting hemodynamics Yet marked improvement in functional capacity possibly due to increased blood perfusion Remains a suitable candidate on the active heart transplant list 9 Confidential © 2014 Sunshine Heart, Inc. |
Implications Caution: Single patient experience Preliminary findings concur with large scale studies with B-blockers1,2 and cardiac resynchronisation therapy pacemakers3 - Subtle improvements in heart function can result in clinically important benefits, i.e. improved exercise tolerance and quality of life. 1.NEJM 1996; 334:1349 2.NEJM 2001;344:1651 3. NEJM 2002;346:1845 10 Confidential © 2014 Sunshine Heart, Inc. |
Exhibit 99.4
Progress of the Fully Implantable System Sep 16, 2014 www.sunshineheart.com |
Internal electro-hydraulic converter and TETS eliminate the percutaneous drive line and associated infection risks. Non-blood contacting Non-obligatory C-Pulse II Overview © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. Fully Implantable System |
Internal electro-hydraulic converter and TETS eliminate the percutaneous drive line and associated infection risks. Non-blood contacting Non-obligatory No percutaneous drive line © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. C-Pulse II Overview Fully Implantable System |
Internal electro-hydraulic converter and TETS eliminate the percutaneous drive line and associated infection risks. Non-blood contacting Non-obligatory No percutaneous drive line No implanted battery © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. C-Pulse II Overview Fully Implantable System |
Pramod Bonde, et al. Yale University © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 4 |
Pramod Bonde et al. Yale University © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 5 |
Eliminating the driveline is not a new idea Circa 1985 © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 6 |
Arrow LionHeart LVAD First clinical TETS system © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 7 |
Arrow LionHeart LVAD First clinical TETS system © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 8 |
Arrow LionHeart LVAD First clinical TETS system © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 9 |
Penn State Arrow Lionheart summary Nov 9, 2003 23 male patients enrolled between Oct 1999 and Dec 2002 10/23 discharged home with device 8/23 alive at 2 years 1/23 alive at 3 years Mean duration 347 days (17-1259) 5/23 serious infections (.17/patient year vs. .60/patient year REMATCH no deaths due to infections) No serious TETs complications © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 10 |
23 male patients enrolled between Oct 1999 and Dec 2002 10/23 discharged home with device 8/23 alive at 2 years 1/23 alive at 3 years Mean duration 347 days (17-1259) 5/23 serious infections (.17/patient year vs. .60/patient year REMATCH no deaths due to infections) No serious TETs complications Penn State Arrow Lionheart summary Nov 9, 2003 © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 11 |
Where will clinical implementation of TETS technology first find traction? © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 12 |
© 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 13 |
© 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. ? 14 |
© 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. SUNSHINE HEART C-pulse Extra-Aortic Balloon Cuff 15 |
SUNSHINE HEART C-pulse Extra-Aortic Balloon Cuff © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. Non-obligatory No blood contact so 16 |
SUNSHINE HEART C-pulse II © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 17 |
Electrohydraulic pump © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 18 |
C-Pulse II Leveraging powerful synergies Minnetronix Cirtec Medical Systems Texas Heart Institute At St. Lukes Episcopal Hospital © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 19 |
Electrohydraulic pump Cirtec Medical Systems Rapidly inflates and deflates the extra-aortic balloon cuff (reproduces C-I physiology) EKG synchronized to provide counter-pulsation Balloon passively empties in the event of pump or power failure (essential) Leverages the incompressibility of silicone oil Compliance reservoir incorporated into the base of the pump © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 20 |
Rapidly inflates and deflates the extra-aortic balloon cuff (reproduces C-I physiology) EKG synchronized to provide counter-pulsation Balloon passively empties in the event of pump or power failure (essential) Leverages the incompressibility of silicone oil Compliance reservoir incorporated into the base of the pump © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. Electrohydraulic pump Cirtec Medical Systems 21 |
C-Pulse II Important IP progress © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 22 |
C-Pulse II Important IP progress So no need for an implantable battery © 2014 Sunshine Heart, Inc. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 23 |
Cirtec Medical Systems Electrohydraulic pump- CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 24 © 2014 Sunshine Heart, Inc. |
How are we going to power it? CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 25 © 2014 Sunshine Heart, Inc. |
Trans-cutaneous Energy Transfer System (TETS) DC battery pack outside the body DC current is put through an oscillator to make AC AC current energizes external coil (1o) to generate an oscillating magnetic field Oscillating magnetic field goes through the skin Oscillating magnetic field is picked up by a tuned internal coil (2o) resulting in induction of AC current AC current rectified into DC used to run the internal device CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 26 © 2014 Sunshine Heart, Inc. |
Standard transformer CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 27 © 2014 Sunshine Heart, Inc. |
Inductive coupling through an air-gap CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. Design minimizes risk of skin heating 28 © 2014 Sunshine Heart, Inc. |
July 10 1856 January 7 1943 Nikola Tesla CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 29 © 2014 Sunshine Heart, Inc. |
Newest systems are: Smaller size so easier to implant More energy efficient so improved battery life More tolerant of geometric misalignment Minnetronix Leaders in Transcutaneous Energy Transfer Systems (TETS) CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 30 © 2014 Sunshine Heart, Inc. |
Minnetronix Leaders in Transcutaneous Energy Transfer Systems (TETS) CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 31 © 2014 Sunshine Heart, Inc. |
Improvement in TETS component geometry and function Minnetronix CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 32 © 2014 Sunshine Heart, Inc. |
Improvement in TETS component geometry and function Minnetronix CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 33 © 2014 Sunshine Heart, Inc. |
How are we going to power it? Where are we going to test it? CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 34 © 2014 Sunshine Heart, Inc. |
CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. TEXAS HEART INSTITUTE at St. Lukes Episcopal Hospital © 2014 Sunshine Heart, Inc. 35 |
THI Cardio Vascular Research Lab ICU 36 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
Surgical Implant and System Integration THIs Cardio Vascular Research Lab The premiere large animal cardiovascular research lab in the world Non-clinical and pre-clinical safety testing are required by the global regulations prior to human trials' Domain dominance in development and implementation of heart failure technology Successful acute system implantation (first generation) CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. 37 © 2014 Sunshine Heart, Inc. |
Minimally Invasive Implantation Small incision Sternal-sparing No cardiopulmonary bypass 38 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
Recent Progress -Bench-Top Tests, CPII versus CPI Typical Average power 3Watts Peak Instantaneous 20W 39 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
40 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
41 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
Recent Acute animals (x2) 42 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
Recent Chronic animal 21 days continuous operation 43 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
44 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |
CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. 45 |
The SUNSHINE HEART C-Pulse II has the potential to be the first completely self-contained therapy for heart failure since the bi-ventricular pacer Lack of blood contact and non-obligatory feature make it the most likely candidate to leverage TETS in a mechanical circulatory assist device Pump innovation has facilitated development of a novel technology, avoiding the safety and regulatory risks of an implantable battery The system is well suited for implantation off-pump through a small sternal sparing incision, making it well suited for patients earlier in the course of heart failure Early pre-clinical testing suggests the design is performing as intended and within established safety parameters In summary 46 CAUTION: Investigational Device. Limited by Federal (or United States) Law to Investigational Use. © 2014 Sunshine Heart, Inc. |